Why are calibration and QC program date- and timestamps incorrect on my KaVo Dental equipment?
- EEdward HouseSep 12, 2025
If the calibration and QC program date- and timestamps are incorrect, it means the device didn't receive the system time correctly during start-up. Here's what you can do: 1. Restart the device and redo a calibration or QC program to check if the device receives the time correctly. 2. Check that the system time is correct on the acquisition workstation, or any other computer that may operate as network time server in the local network. 3. Ensure that the acquisition workstation has UDP port 123 opened in the firewall and the NetTime service installed and running.