Questions and Answers

  • S
    Samuel HarrisSep 12, 2025
    Why is there no heating function on the multifunctional handpiece of my KaVo Medical Equipment?
    • K
      Kristen HallSep 12, 2025
      If there's no heating function on the multifunctional handpiece of your KaVo Medical Equipment, it's possible that spray heating is not preselected. Pre-select spray heating.
  • D
    David CarterSep 18, 2025
    What does it mean if the LED on the "AP1" button flickers on my KaVo Medical Equipment?
    • R
      rachel99Sep 18, 2025
      If the LED on the "AP1" button flickers on your KaVo Medical Equipment's dentist element, the data connection to the assistant element is faulty.
  • O
    ofloydSep 4, 2025
    What to do if the ERGOcam does not work with my KaVo Medical Equipment?
    • J
      Justin JohnstonSep 4, 2025
      If the ERGOcam isn't working with your KaVo Medical Equipment, check if the PC is switched off and turn it on.
  • C
    Christopher WarnerAug 26, 2025
    What to do if the display shows CAN fail on my KaVo Medical Equipment?
    • S
      stephenstamaraAug 27, 2025
      If the display on your KaVo Medical Equipment shows 'CAN fail', it indicates an internal communication error. Turn the unit off and on again.
  • U
    urodgersAug 27, 2025
    What to do if nothing works on my KaVo Medical Equipment?
    • A
      andreachaseAug 27, 2025
      If your KaVo Medical Equipment isn't working, it might be due to the main switch being off. Turn the main switch on. Alternatively, the main service fuse may have interrupted the electric circuit. Unplug the unit from the mains, then check and replace the main service fuse if needed.
  • S
    Sandra TurnerAug 28, 2025
    What to do if KaVo Medical Equipment suction hoses do not have any suction?
    • A
      amy66Aug 28, 2025
      If the suction hoses on your KaVo Medical Equipment aren't working, check if the sliders on the conical sections are closed and open them. Also, the sieves in the suction connector could be clogged, so replace them. Finally, make sure the chair kick plate is not being actuated; release it if it is.
  • W
    William MejiaSep 22, 2025
    What to do if several handpieces are simultaneously activated on my KaVo Medical Equipment?
    • M
      Megan EstesSep 22, 2025
      If several handpieces are simultaneously activated on your KaVo Medical Equipment, it indicates a hardware error. Stop working.
  • L
    Leslie WatsonSep 23, 2025
    What to do if an acoustic signal is issued every second on my KaVo Medical Equipment?
    • M
      Melissa GardnerSep 23, 2025
      If an acoustic signal is issued every second on your KaVo Medical Equipment, it means a leaking water switch recognizes leaking water. Remove water from the unit body.
  • D
    diazjillSep 12, 2025
    What does it mean if beeps every 10 seconds and the LED on the "Intensive germ reduction" key flashes (green) on my KaVo Medical Equipment?
    • M
      Martin HaysSep 12, 2025
      If your KaVo Medical Equipment beeps every 10 seconds and the LED on the "Intensive germ reduction" key flashes green, the Oxygenal container is likely empty. Refill the Oxygenal container.
  • T
    Tammy FitzgeraldSep 4, 2025
    What to do if the LED on the "HYDROclean" button (red) flashes on my KaVo Medical Equipment?
    • L
      Laurie CarterSep 5, 2025
      If the LED on the "HYDROclean" button (red) flashes on your KaVo Medical Equipment, it indicates a malfunction in the amalgam separator.

Summary

Safety Guidelines

Safety Instruction Structure and Hazard Levels

Explains warning symbols, structure, and different hazard levels.

Purpose and Proper Use Guidelines

Defines intended use, proper operation, and product-specific usage.

General and Product-Specific Safety Instructions

Covers general safety, explosion hazards, and specific risks.

Product Description

Operation Guide

Adjusting the Dental Chair

Detailed guide on adjusting chair positions, headrests, and safety features.

Using Menu Functions

Guide to navigating menus for various settings like firmware, time, language, and MEMOspeed.

Using Dentist and Assistant Unit Functions

Details on hygiene, light, bell, timer, and saving handpiece settings.

Operating the Foot Switch

Explains general functions, chair positioning, and instrument control via foot switch.

Safety Check - Test Instructions

Check Components and Intervals

Details inspection types, test methods, and testing intervals.

Measurements and Functional Tests

Guides on performing measurements, functional tests, and assessment/documentation.

Troubleshooting

Malfunctions and Remedies

Lists common malfunctions and their causes and remedies.

Electromagnetic Compatibility (EMC) Data

KaVo Primus 1058 Life TM Specifications

General IconGeneral
BrandKaVo
ModelPrimus 1058 Life TM
CategoryMedical Equipment
LanguageEnglish

Related product manuals