Series 3700A System Switch/Multimeter Reference Manual Section 11: TSP command reference
3700AS-901-01 Rev. D/June 2018 11-151
An error is generated if:
• You try to set dmm.autozero for "continuity" or "nofunction". Error code 1114, "Setting
conflicts with function selected," is generated.
• You query dmm.autozero for "continuity" or "nofunction". nil is returned with error
code 1114, "Setting conflicts with function selected."
Example
dmm.func = "dcvolts"
dmm.autozero = dmm.ON
dmm.autozero = dmm.AUTOZERO_ONCE
print(dmm.autozero)
timer.reset()
dmm.autozero=2
time=timer.measure.t()
print(time)
Enables autozero for DC volts.
Refreshes the reference points once and sets
autozero to dmm.OFF or 0.
Output:
0.000000000e+00
Determines the time delay when autozero is selected.
Also see
Autozero (on page 4-3)
dmm.configure.recall() (on page 11-165)
dmm.configure.set() (on page 11-166)
dmm.func (on page 11-179)
dmm.nplc (on page 11-209)
dmm.reset() (on page 11-219)
reset() (on page 11-308)