B-6 General Measurement Considerations Model 6485 and 6487 User’s Manual
The effect on instrument performance can be considerable if enough of the unwanted sig-
nal is present. The effects of EMI can be seen as
an unusually large offset or, in the case of
impulse sources, erratic variations in the displayed reading.
The instrument and experiment should be kept
as far away as possible from any EMI
sources. Additional shielding of the instrument, experiment and test leads will often
reduce EMI to an acceptable level. In extreme cases, a specially constructed screen room
may be required to sufficiently attenuate the troublesome signal.
External filtering of the input signal path may be required. In some cases, a simple one-pole
fi
lter may be sufficient. In more difficult situations, multiple notch or band-stop filters, tuned
to the offending frequency range, may be required. Connecting multiple capacitors of widely
different values in parallel will maintain a low impedance across a wide frequency range.
Keep in mind, however, that such filtering may have detrimental effects (such as increased
response time) on the measurement.

Table of Contents

Other manuals for Keithley 6485

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 6485 and is the answer not in the manual?

Keithley 6485 Specifications

General IconGeneral
ModelKeithley 6485
CategoryPicoammeter
Input Impedance>10^14 Ohms
Voltage Burden<200µV
Accuracy0.4% of reading
Voltage Source RangeNot Applicable
Voltage Source ResolutionNot Applicable
Voltage Source AccuracyNot Applicable
InterfaceIEEE-488
Power Requirements100V to 240V AC, 50/60Hz
Dimensions89 mm high x 214 mm wide x 369 mm deep (3.5 in x 8.4 in x 14.5 in)

Related product manuals