Section
5: Measuring low-resistance devices Model 2460 Interactive SourceMeterĀ® Instrument
User's Manual
5-8 2460-900-01 Rev. A / September 2014
Send the following commands for this example application:
--Reset the instrument to the default settings
reset()
--Configure the Simple Loop trigger model template to make 100 readings.
trigger.model.load("SimpleLoop", 100)
--Change the view on the front panel to the TREND swipe screen.
display.changescreen(display.SCREEN_PLOT_SWIPE)
--Set to measure resistance, use 4-wire sense,
--and offset compensation.
smu.measure.func = smu.FUNC_RESISTANCE
smu.measure.sense = smu.SENSE_4WIRE
smu.measure.offsetcompensation = smu.ON
--Turn on the output
smu.source.output = smu.ON
--Initiate trigger model and wait until finished.
trigger.model.initiate()
waitcomplete()
--Turn off output
smu.source.output = smu.OFF
--Read the resistance and time values from defbuffer1.
print("Resistance:\tTime:")
for i = 1, 100 do
print(string.format("%f\t%f", defbuffer1[i], defbuffer1.relativetimestamps[i]))
end

Table of Contents

Other manuals for Keithley SourceMeter 2460

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley SourceMeter 2460 and is the answer not in the manual?

Keithley SourceMeter 2460 Specifications

General IconGeneral
BrandKeithley
ModelSourceMeter 2460
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals