12-106 Instrument Control Library Series 2600 System SourceMeters Reference Manual
2600S-901-01 Rev. A / May 2006 Return to Section 12 topics
status.node_event
Attribute Status node event register.
Usage Reads the status node event register:
nodeeventreg = status.node_event
Remarks
This attribute is used to read the status node event register, which is returned as a
numeric value. Reading this register returns a value. The binary equivalent of the
returned value. The least significant bit of the binary number is bit 0, and the most
significant bit is bit 7.
For example, assume value 129 is returned for the event register. The binary
equivalent is 10000001. This value indicates that bit B0 (MSB) and bit B7 (OSB) are
set.
The status node event register uses most of the same summary events as the status
byte. Bit B1(MSB) is not used, and bit B6 is used as Master Summary Status (MSS).
For details, see “
status.condition” register.
Details See “Status byte and service request (SRQ)” in Appendix D.
Example Reads the status node event register:
nodeeventreg = status.node_event
print(nodeeventreg)
Output: 1.29000e+02
The above output indicates that bits B0 (MSS) and B7 (OSB) are set.

Table of Contents

Other manuals for Keithley SourceMeter 2602

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley SourceMeter 2602 and is the answer not in the manual?

Keithley SourceMeter 2602 Specifications

General IconGeneral
BrandKeithley
ModelSourceMeter 2602
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals