13-22 Factory Scripts Series 2600 System SourceMeters Reference Manual
2600S-901-01 Rev. A / May 2006 Return to Section 13 topics
Waveform:
Pulse initiated here
bias
level
toff
start
Measurement (1/NPLC seconds)
step
stop
ton
step = (stop - start)/(points -1)
Last source value before pulse train was initiated
step
Return
Values:
f: A Boolean flag. This flag will be true when the pulse was
successfully configured, false when errors were encountered.
msg: A string message. If the f flag is false, msg will contain an error
message. Otherwise, msg will contain a string indicating successful
configuration.
Output
Data:
Pulsed current measurements are stored in the reading buffer specified by
the buffer input argument.
Example:
ConfigPulseVMeasureISweepLin(smub, 0, 1, 10, 1, 10e-3,
20e-3, 16, smub.nvbuffer1, 4)
Setup a pulsed sweep that will use SourceMeter channel B. The pulsed
sweep will start at 1V, end at 10V, and return to a 0V bias level between
pulses. Each pulsed-step will be on for 10msec and then at the bias level for
20msec. The current limit will be 1A during the entire pulsed-sweep. The
pulse train will be comprised of 16 pulsed-steps and the pulse train will be
assigned a tag index of 4.
See Also: InitiatePulseTest, InitiatePulseTestDual