12-4 Instrument Control Library Series 2600 System SourceMeters Reference Manual
2600S-901-01 Rev. A / May 2006 Return to Section 12 topics
Either of the following command messages will configure SMU A for the moving
average filter:
smua.measure.filter.type = 0
smua.measure.filter.type = smua.FILTER_MOVING_AVG
Some attributes can take any numeric value that is within a valid range. For
example, the Model 2601/2602 voltage source can be set from -40.4V to +40.4V,
while the Model 2611/2612 voltage source can be set from -202V to +202V. The
following command message sets the
SMU A source level to 1.53V:
smua.source.levelv = 1.53
Reading an attribute
Reading an attribute is accomplished by passing it to a function call as a
parameter or by assigning it to another variable.
Parameter passing example – The following command reads the filter type for
SMU A by passing the attribute to the
print function, which outputs a value:
print(smua.measure.filter.type)
Output: 0.000000e+00
The above output indicates that the moving average filter is selected.
Variable assignment example – The following command reads the filter type by
assigning the attribute to a variable named
filtertype:
filtertype = smua.measure.filter.type
Syntax rules
• Commands for functions and attributes are case sensitive. As a general
rule, all function and attribute names must be in lower case, while
parameters use a combination of lower and upper case characters. Upper
case characters are required for attribute constants. Example:
smua.source.func = smua.OUTPUT_DCVOLTS
In the above command to select the volts source function, OUTPUT_DCVOLTS
is the attribute constant.
• Whitespace in a function is not required. The function to set digital I/O line 3
low can be sent with or without whitespaces as follows:
digio.writebit(3,0) -- Whitespaces NOT used in string.
digio.writebit (3, 0) -- Whitespaces used in string.
• Some commands require multiple parameters. Multiple parameters must
be separated by commas (,), as shown above for the
digio.writebit
function.

Table of Contents

Other manuals for Keithley SourceMeter 2611

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley SourceMeter 2611 and is the answer not in the manual?

Keithley SourceMeter 2611 Specifications

General IconGeneral
BrandKeithley
ModelSourceMeter 2611
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals