KEOFITT W9 USER MANUAL V.4 PAGE 25
1. Open the valve slowly and take the sample.
NB! In order to ush the valve seat before sampling you may choose
to let the rst few seconds of ow go to drain.
2. Close the valve aer the sample has been taken.
3. Clean the valve by ushing with steam, water or hot water.
If the process media is sticky, viscous or aggressive or for any other appropriate reason, do repeat a full
CIP cycle aer sampling in case steam is not available and ushing with water prove insuicient.
WARNING
• When sampling at a high pressure and/or with a low viscosity process media it may ow rapidly
into the sample recipient. Therefore open the valve slowly. Special care must be taken with
pneumatically operated valves, as they open abruptly. If problematic, consider adjustable
actuators.
• Always remember to wear safety goggles when steaming, CIPping, taking samples or any other
operations of the sampling valve