EasyManua.ls Logo

KEP BATCHtrol II - Page 28

KEP BATCHtrol II
51 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
24
3-10 Setup Procedure For The Batcher
3-10 Setup Procedure For The Batcher
(continued)
From the previous page, we are in the
SET HIGH
setting portion of the ALG OUT setup Menu.
PRESS
DISPLAY
0 Flashes.
Clears out existing High Setpoint value.
6759.5 Flashes.
Sample High Setpoint, or enter value from notes or worksheet. (D for decimal point.)
If High Setpoint is too low, the warning HIGH
LOW will be displayed and the unit will
return to the SET LOW routine.
Last count, unit now in Run Mode.
High Setpoint is stored.
Analog Output is set.
MENU ITEM 9
OUT FREQ
All models of the Batcher have a pulse generator built in to them.
This Output Frequency
generator sends pulses out which are scaled relative to the input signal via the
counter
K
Factor.
This means that for every time the counter increments a pulse is available at the output.
Various output frequencies are available to the user for driving external devices. In case the rate
exceeds the output frequency selected, a 9999 pulse buffer is provided to hold the excess pulses.
If the buffer is completely fi lled the warning message DATALOST will fl ash on the display.
PRESS
DISPLAY
PRESET
D PREWARN
Press until . . . OUT FREQ
20000
Enters the routine for setting up the Frequency Output.
The display shows the last Frequency selection.
2000
200
10
Press D to go to 20000
Press ENT at desired Frequency.
Last count unit now in Run Mode.
Pulse Output is now set.
CLR
ENT
6 7 5 9 D 5
D
D
D
ENT
D
D
D
ENT