EasyManuals Logo
Home>KERN>Scales>ALJ Series

KERN ALJ Series User Manual

KERN ALJ Series
Go to English
233 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #112 background imageLoading...
Page #112 background image
ALJ/ALS/PLJ/PLS-BA-e-1640
34
7.2.2 Overwrite internal adjustment weight
(Models with settings not appropriate for verification only)
! Overwriting is restricted to specialist staff possessing well acquainted with the
workings of weighing scales.
Press the MENU button and keep it pressed until the acoustic signal gets mute.
Press the MENU button as often as required until „CAlib“ appears
Operate the PRINT key
Press the MENU button as often as required until „TEC-CAL appears
Press the PRINT button and keep it pressed until the acoustic signal gets mute.
Press the MENU button and keep it pressed until the acoustic signal gets mute.
The
balance changes automatically into weighing mode
Ensure that there is no load on the weighing plate
Press the CAL key
Wait until the exact value of the adjustment weight will be shown flashing
(Example)
Place the displayed adjusting weight on the weighing plate.
The flashing display extinguishes and the balance changes into weighing mode.
Remove adjustment weight from weighing plate
Press the PRINT button and keep it pressed until the acoustic signal gets mute.
The calibrating process is started
During this process „tEc MEM“ is displayed
After having automatically saved the value of the internal adjustment weight, the balance
will return into weighing mode.
Carry out the adjustment process as described in chapter 7.2

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the KERN ALJ Series and is the answer not in the manual?

KERN ALJ Series Specifications

General IconGeneral
BrandKERN
ModelALJ Series
CategoryScales
LanguageEnglish

Related product manuals