EasyManuals Logo

KERN ALJ Series User Manual

KERN ALJ Series
Go to English
233 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #122 background imageLoading...
Page #122 background image
ALJ/ALS/PLJ/PLS-BA-e-1640
44
9.2 Data output type
In weighing mode press the MENU button and keep it pressed until the acoustic
signal gets mute.
Press MENU button
Acknowledge using PRINT button, the current setting is displayed.
Press the arrow keys to select desired setting.
Scroll forward using MENU button
Scroll backward using CAL button
„PC cont“ = Continuous output
„PC CMd“ = Manual issue after pressing the PRINT button
„Print“ = not documented
Take over selection using the PRINT button.
Weighing balance returns to menu. Either make more settings in the menu or go
back to weighing mode as follows.
Press the MENU button and keep it pressed until the acoustic signal gets mute.
The balance returns automatically into weighing mode.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the KERN ALJ Series and is the answer not in the manual?

KERN ALJ Series Specifications

General IconGeneral
BrandKERN
ModelALJ Series
CategoryScales
LanguageEnglish

Related product manuals