9.5 Filter
This menu item allows the balance to be set according to specific ambient conditions
and measuring purposes.
In weighing mode press the MENU button and keep it pressed until the acoustic
signal gets mute.
Press MENU button repeatedly
Acknowledge using PRINT button, the current setting is displayed.
Press the arrow keys to select desired setting.
Scroll forward using MENU button
Scroll backward using CAL button
FILTER 1: Setting for dispensing
FILTER 2: Sensitive and fast, very quiet set-up location.
FILTER 3: Robust but slow, busy set-up location.
Take over selection using the PRINT button.
Weighing balance returns to menu. Either make more settings in the menu or go
back to weighing mode as follows.
Press the MENU button and keep it pressed until the acoustic signal gets mute.
The balance returns automatically into weighing mode.