ALJ/ALS/PLJ/PLS-BA-e-1640 47
Press the arrow keys to select desired setting.
Scroll forward using MENU button
Scroll backward using CAL button
Au0 OFF = Auto Zero switched off
Au0 1 = Auto Zero range ± ½ digit
Au0 2 = Auto Zero range ± 3 digits
Au0 3 = Auto Zero range ± 7 digits
Au0 3E = Auto Zero range ± 7 digits in the whole weighing range
Take over selection using the PRINT button.
Weighing balance returns to menu. Either make more settings in the menu or go
back to weighing mode as follows.
Press the MENU button and keep it pressed until the acoustic signal gets mute.
The balance returns automatically into weighing mode.
For models PLJ_M settings are restricted to “Au0 OFF“ and „Au0 1“