8.4.3 Dosing and Zero-tracking
The Auto-Zero function is used to tare small variations in weight automatically.
In the event that small quantities are removed or added to the material to be
weighed, incorrect weighing results can be displayed due to the “stability
compensation“ in the balance. (Example: Slowly draining fluids from a container on
the balance.)
When apportioning involves small variations of weight, it is advisable to switch off this
function.
If Zero-Tracking however is switched off, the weighing display becomes more busy.
Ö With the weighing scale switched off, press ON/OFF and TARE-key
simultaneously until the first function “Unit“ appears.
Ö Repeatedly press the 10x-key until “tr“ appears.
Ö Confirm by pressing the PRINT-key. Current setting appears.
Ö Now you can use the 10x-key to select among the settings below:
„tr on“: Function activated
„tr off“: Function deactivated
Ö Changed setting will be imported by pressing the PRINT key. The balance
returns automatically into weighing mode.