EasyManuals Logo

KERN CME 100-3 User Manual

KERN CME 100-3
35 pages
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #35 background imageLoading...
Page #35 background image
CDE/CME/CDEE-BA-e-1014
35
12.4 Installation
Switch off both scales and disconnect
Connect both scales with the data cable provided for the RS232 interface
12.5 Balance settings
To enable automatic transfer of the reference weight from CME to CDE, the data
transfer mode “Pr” in the menu (see chpt.9.3.1) must be set to “CSyS“.
12.6 Counting with both scales
Attention:
To prevent errors during the piece count determination, both scales must be adjusted
to the same gravity acceleration (see chap. 6.8).
Non-compliance will result in counting errors!
1. Place empty vessel on weighing platform of quantity scale KERN CDE and tare
by pressing the key (TARE)
2. Place reference piece number of count quantity on reference scale KERN CME
(e.g. 5, 10, 20 items)
3. Confirm selected reference piece number by pressing the key (5, 10, 20) on the
reference scale.
4. Fill vessel on quantity scale with count quantity. The piece number is shown
directly on the display. The -key may be used to switch over to piece weight.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the KERN CME 100-3 and is the answer not in the manual?

KERN CME 100-3 Specifications

General IconGeneral
BrandKERN
ModelCME 100-3
CategoryScales
LanguageEnglish

Related product manuals