p Put the 1
st
weight mass (e.g. weight mass=3kg, *n=1), then press . key
Weight Piece Weight Pcs
0 2 1 XXXXXX
input *n value internal value of the phase
q Add the 2
nd
weight mass (e.g. 6kg, in sum 9kg) and input *n value (e.g. *n=2),
then press . key
Weight Piece Weight Pcs
0 3 1 XXXXXX
input *n value, e.g. n=2 internal value of the phase
r Add the 3
rd
weight mass (e.g. 12kg, in sum 21kg) and input *n value
(e.g. *n=4), then press . key
Weight Piece Weight Pcs
0 4 1 XXXXXX
input *n value, e.g. n=4 internal value of the phase
s Add the 4
th
weight mass (e.g. 12kg, in sum 33kg) and input *n value
(e.g. *n=4), then press
. key
Weight Piece Weight Pcs
0 5 1 XXXXXX
input *n value, e.g. n=4 internal value of the phase
CXB-SH-e-0913 13