4 Error messages
Non-approval models:
E1 ZERO over 300,000 internal value.
E2 ZERO below 10,000 internal value.
Approval models:
E1 ZERO over the zero position in weight calibration + 10% of full scale.
E2 ZERO over the zero position in weight calibration - 10% of full scale.
All models:
E4 ZERO is unstable.
E5 Turning on machine core value is under 0.
E6 Core value is over 250,000.
E7 Core value is below 120,000.
- - OL - The weighing range is exceeded.
NOTES:
Non-approval models (CXB): After E1, E2 or E4 is displayed press
to continue.
Approval models (CXB-M): After E1, E2 or E4 appear, set the jumper SWA1
on main board to “ADJ” position and press to return the counting
mode.
Then press
+ + at the same time followed by pressing
to access calibration (see chap. 7.3).