11 Temperature calibration
♦ Before calibrating, please adjust the mini jumper SWA1 to “ADJ” position.
♦ After finish the calibration, please re-adjust the mini jumper SWA1 to “LOCK”
position.
11.1 Indoor air (normal) temperature calibration
n * The scale will be self-tested and reset back to zero.
Press key+ 1 key+ 5 key, then press key to enter the temperature
calibration mode.
Weight Piece Weight Pcs
0. 0 0 0 XX XXXXXX
indoor temperature(°C) zero internal value
o No weight on the platter, press . key to calibrate zero.
Weight Piece Weight Pcs
3 0. 0 0 0 XX XXXXXX
full capacity value indoor temperature(°C) internal value of full capacity
p Put the full capacity weight on the platter, press . key to calibrate full capacity
value.
q After scale is stable, it will reset back to zero automatically (remove the weight
mass from the platter before it resets back to zero), then the procedures
finished. Put the weight mass on the platter again to check if the display
shows the accurate weight. If it’s correct, turn off the power, then re-adjust the
mini jumper SWA1 to “LOCK” position.
NOTES:
(1) The procedures are for the scale without the temperature IC or for the scale in
the normal temperature environment.
(2) To achieve the accuracy of the scale, please run the high temperature and low
temperature calibration procedures.
(3) The procedures for high temperature and low temperature are totally reversed.
18 CXB-SH-e-0913