e different versions of models setting
0 = Non-approval model (External calibration is applicable)
1 = OIML approval model (The display resolution should be set at 1/3000,
External calibration is NOT applicable)
4 = Non-approval model (External calibration is NOT applicable)
p After input spec., press
key (not necessary to put the weight mass on the
platter, the calibrated weight value is the reference for the temperature
calibration procedure).
The scale will be self-tested and reset back to zero.
Press
key 1 key 5 key, then press key to enter the temperature
calibration mode (see chapter 11).
NOTES:
(1) The last number of full capacity value must be variable number, can be fixed “0”.
(2) The digit and location of preset calibrated value, refer to the location of capacity
digit.
(3) Input capacity + 9d in order to run 9 times overload function.
(4) If max. number of digit is “0”, please input “0”.
(5) Reference values for capacity setting see chapter 10.1.