3 Calibration Procedure (CAL)
1. Turn scale on by pressing the [ON/TARE] key.
2. Press the [OFF] key for approx. 5 seconds. Display will show [CAL] and then the
required calibration weight.
3. Gently place the correct calibration weight on the center on the weighing pan.
Wait until display shows [ F ] then turn off.
In case [ E] is display instead of [ F], this indicates a calibration procedure error,
unstable or wrong calibration weight applied for calibration.
Turn the scale off and then on and repeat the procedure.