7 Trouble Shooting
Power on
Full Segments?
If no display, check battery /adaptor, keys, connection between
battery/adaptor—main board.
If missing segments, check fixing of LCD frame, zebra connector under LCD.
Display Zero?
If display [LO], check battery >2.6v
(>7.5v for EMB 100-3,EMB 200-2,EMB 600-2,EMB 500-1,EMB 1000-2,
EMB 1200-1,EMB 5.2K1)
adaptor>9v.
If display [ E] , check internal count.
Proper readout?
If unstable reading, check weighing plate, overload stopper, load cell and wires,
environmental conditions and stable table.
Correct reading?
If not accurate, perform internal calibration.
If cannot reach full capacity, check weighing plate, overload stopper, load cell
and internal zero point.
If always zero, check internal zero point. Internal calibration if necessary.
Normal operation.