6.7.2 Automatic internal adjustment
Automatic adjustment takes place:
when the balance is switched off and on.
at the end of a time interval
After a set time interval (selectable 1 8 hrs.) in the menu (F5 HoUr, see chap.
8.3) the internal adjustment is automatically started.
The motor noise of the loading system for the internal adjustment weight can be
heard, the internal adjustment is started.
After successful adjustment „PASS“ will be displayed. The balance returns
automatically into weighing mode.