FTC-BA-e-0720
29
Settings in the menu
User manual ME-Nr.22017145 01/07
by an accredited organization, and a new certification seal must be affixed before
the instrument may be used as a certified scale again!
• Hold down the service switch (pushbutton) on the analogprint (indicated by an
arrow in the drawings below, while at the same time switching on the scale. Keep
the service switch pressed in until “Service” appears in the display.
Small platform model Large platform model
1.Display "Scale": Press the -key (within 20s)
2.Display "Metrolo": Press the -key
3.Display "Scale 1": Press the -key
4.Display "ramp": Press the -Tkey
5.Display "SNR": Press the -key
6.Display "SCAL.bld": Press the -key
7.Display "GEO" (adjustment by means of GEO value):
The GEO value can be set in this block and the balance also adapted to the local grav-
ity ratios without adjusting weights.
Case a)You are familiar with the GEO values.
Once the -key has been pressed, the current GEO value will be displayed.
Press the oder -key, in order to alter the GEO value. The next value is dis-
played every time the key is pressed (adjustment range: 0 - 31). See the GEO value
chart in chapter 7.1.1 for the appropriate value.
Confirm the selected GEO value using the -key.
Please note: The GEO value may not be readjusted following this "adjustment by
means of GEO value", as this would cause the set adjusting values to become invalid.
Fall b) Sie sind mit den GEO-Werten NICHT vertraut. Die Justierung muss in diesem
Fall mit Justiergewicht erfolgen (siehe Punkt 8).
You are NOT familiar with the GEO values. In this case adjustment must be made
using adjusting weights (see item 9).
- Taste drücken