Settings in the menuFTC-BA-e-0720
30User manual ME-Nr.22017145 01/07
7.Display "LIN-CAL": Press -key
8.Display "CAL": Press -key
Press the -key.The scale determines the zero point. -preload- appears in the
display. The calibration weight to be placed on the scale then flashes in the display.
If necessary, change the weight value displayed with .
Place the calibration weight on the scale and confirm with .
The scale calibrates with the calibration weight loaded. After calibration is completed,
-done- appeats briefly in the display.
Return to weighing mode:
Press the -key, "END" will appear on the display :
Press the -key, order to save the alterations. "Save" will appear on the display.
Press the -key to confirm. Following this the balance will return to weighing
mode.
4.3.2 DISPLAY – weighing unit and display accuracy
4.3.3 TARE – tare function
4.3.4 ZERO – automatic zero update
UNIt1 Select weighing unit 1: g, kg, oz, lb, t
UNIt2 Select weighing unit 2: g, kg, oz, lb, t
rESOLU Select readability (resolution), model-dependent
UNt.rOLL When UNT.rOLL is switched on, the weight value can be displayed in all available
units and as pieces with .
Notes
• On certified scales, the weighing units oz and lb are displayed with the symbol *.
• On certified scales, resolutions that deviate from the scale definition are displayed
without a weighing unit and with the symbol *.
• On dual-range/dual interval scales, resolutions marked with |<–> 1/2| are
divided up into 2 weighing ranges / intervals, e.g. 2 x 3000 d.
A-tArE Switching on/off automatic taring
CHAIn.tr Switching on/off chain tare
A.CL-tr Switching on/off automatic taring with automatic clearing of the tare weight when the
load is removed from scale
AZM On certified scales, this menu item does not appear.
Switching on/off automatic zero update and selecting zeroing range.
Possible settings: OFF; 0.5 d; 1 d; 2 d; 5 d; 10 d