5.1 Data Printout/Record (ISO/GLP-compliant)
To have a data record printed out, perform the following:
- Press
PRINT
to output the printout heading and the first value (after you have turned
on the balance or cleared a function by pressing
CF
)
- Press
PRINT
to output additional data
- To end printout generation and recording of data, press
CF
.
Generation of an ISO/GLP-compliant printout or record is also ended when an adjusting
operation is started.
The printout can consist of the following lines:
-----------------
Model 770
: Balance model
S/N 040500046
: Serial number of the balance
Id
: Space for entering a workstation or operator ID
-----------------
Date : 30-May-94
: Current date
Start: 10:05:30
: Start of application
Ser. :
: Space for entering a project number
N + 4.45390g
: Measured weights
N +14.34586g
N +53.23450g
End : 10:05:45
: End of application
Name:
: Space for signature of the operator responsible
-----------------