FTx/ITx-SH-e-0720 55
7.2 Technician menu – Overview
The overview below shows the menu blocks holding additional settings dedicated to the technician (“SCALE”, “TERMINAL”).
The settings available to the technician only are printed in boldface. The further menu blocks (“APPLICATION”,
“COMMUNICATION” and “DIAGNOSTICS”) are similar to the ones in the supervisor setup (see User Manual) and are not shown
below. The factory settings are marked with an asterisk (*). Depending on the model some menu items are not available.
Level 1 Level 2 Level 3 Level 4
SCALE Metrolo no Appr
OIML
NTEP
Scale 1/Scale 2 [choose]
Ramp
Serial Number
Scale Build Scale Type Single Range
2Multi Interval
2Multi Range
3Multi Interval
3Multi Range
Base Unit g, kg, oz, lb, t
Ver. Interval e = d, e = 10d
Scale Cap. 1 [entry]
Resolution 1 1200d ...
Scale Cap. 2 [entry]
Resolution 2 1200d ...
Scale Cap. 3 [entry]
Resolution 3 1200d ...
Geo Value 0 ... 31
LIN-CAL 3 Point [Start]
5 Point [Start]
Calibration [Start]
Control Mode
Display Unit 1 g, kg*, oz, lb, t
Unit 2 g*, kg, t, lb, oz
Resolution 1200d ...
Unt.roll Off*, On
Tare Auto tare Off*, On
Chain tare Off, On*
Auto clear tare Off*, On
Tare key Off, On*
Zero Zero-Capt. –2%+18%*, –2+2%
Set Zero [Start]
Auto zero Off, 0.5d, 1d, 2d, 5d, 10d
Restart Off*, On
Filter Vibration Low, Medium*, High
Process Universal*, Dosing
Stability Fast, Standard*, Precise
Fact Temp Off, 1K, 2K, 5K
MinWeigh Set Value [entry]
On/Off Off*, On
Reset [Start]
TERMINAL Device Language English, US English, German,
French, Italian, Spanish
Power off Off, 1 min., 3 min., 5 min., 15 min., 30 min
Power supply (akku) Off, 1 min., 3 min., 5 min., 15 min., 30 min
Contrast 0 ... 5* ... 10
Weight display small*, large
Date/Time Date format EU, US
Date [entry]
Time [entry]
AM*, PM
Beeper Off*, On
Password Supervisor Password [Enter Code], [Retype Code]
Timeout Mode Not active*, Leasing, Text only
Days 1 ... 999
Text1 [entry]
Text2 [entry]
Reset [Start]