7.3 Assembly and installation
The correct location is a key factor in the accuracy of the weighing results.
Levelling
Exact alignment and stable installation are an essential
requirement for repeatable results. You can
compensate for minor unevenness or tilts of the
footprint by levelling the weighing balance.
Level balance with foot screws until the air bubble of the
water balance is in the prescribed circle.
Check levelling regularly.
Installation of weighing plate
Connection to the power supply (USB)
using the enclosed power pack or the USB host (e.g. computer)
In order to obtain exact results with the electronic balances, the digital
weighing platform must have reached the operating temperature (see warming
The digital weighing platform must be connected to the power
supply during this warming up period.
The accuracy of the digital weighing plat
form is dependent on the local
acceleration due to gravity.
Strictly observe hints in chapter Adjustment.
Connection to your system (configuration) as described in the chapter 8
Carry out adjustment as described in the chapter 0
Communication and using the balance via the interface protocol (see
interface protocol description)
The chapter 10.1 describes the ideal way to use the balance using the
BalanceConnection software.