27 PBS/PBJ-BA-e-1718
If „I-CAL“ is not displayed, press to return into weighing mode and
activate menu element 1, see menu overview.
2. Press the button and adjustment will take place
automatically.
3. After successful adjustment the balance automatically
returns to weighing mode.
In case of an adjustment error (e.g. objects on the
weighing plate) the display will show an error message,
repeat adjustment.
When an optional printer is connected and the GLP
function is connected, the adjustment log will be edited,
see chap. 8.5
Printout example (KERN YKB-01N):
- - - - - - - - - - - - - - - - -
CAL INTERNAL
- - - - - - - - - - - - - - - - -
Adjustment mode
KERN
Company
TYPE
PBJ4200-2M
Model
SN
WBxxxxxxxxx
Serial no.
ID
1234
Balance identification no. (see chap.
8.5.1)
DATE
27-01-2011
TIME
11.54.53
REF
4000.00g
Used adjustment weight
BFR
4003.97g
Before adjustment
AFT
4000.00g
After adjustment
-COMPLETE
-SIGNATURE-
prepared by
- - - - - - - - - - - - - - - - -