16 PFB-BA-e-1322
he adjustment process has been
finished successfully.
Remove adjusting weight during selftest
Wait until the balance is again in weighing mode.
6.8 Linearization
Linearity shows the greatest deviation of a weight display on the scale to the value of
the respective test weight according to plus and minus over the entire weighing
range.
If linearity deviation is discovered during a testing instrument control, you can
improve this by means of linearization.
In balances with a resolution of > 15 000 dividing steps carrying out a
linearization is recommended.
Carrying out linearization is restricted to specialist staff possessing well
acquainted with the workings of weighing scales.
specific
Observe stable environmental conditions. Stabilisation requires a certain
warm-up time.
After successful linearization you will have to carry out calibration;