Loading...
4 - Language Dictionary
78
TRIGger:SEQuence2:SLOPe:CURRent
TRIGger:ACQuire:SLOPe:CURRent
Keysight 66312A, 66332A Only
This command sets the slope of an internally triggered current measurement.
POSitive
triggering occurs on the rising edge.
NEGative
triggering occurs on the falling edge.
EITHer
triggering occurs on either edge.
Command Syntax
TRIGger:SEQuence2:SLOPe:CURRent<slope>
TRIGger:ACQuire:SLOPe:CURRent<slope>
Parameters
EITHer|POSitive|NEGative
*RST Value
EITHer
Examples
TRIG:SEQ2:SLOP:CURR POS TRIG:ACQ:SLOP:CURR EITH
Query Syntax
TRIGger:SEQuence2:SLOPe:CURRent?
TRIGger:ACQuire:SLOPe:CURRent?
Returned Parameters
<CRD>
Related Commands
TRIG:SEQ2:SLOP:VOLT
TRIGger:SEQuence2:SLOPe:VOLTage
TRIGger:ACQuire:SLOPe:VOLTage
Keysight 66312A, 66332A Only
This command sets the slope of an internally triggered voltage measurement.
POSitive
triggering occurs on the rising edge.
NEGative
triggering occurs on the falling edge.
EITHer
triggering occurs on either edge.
Command Syntax
TRIGger:SEQuence2:SLOPe:VOLTage<slope>
TRIGger:ACQuire:SLOPe:VOLTage<slope>
Parameters
EITHer|POSitive|NEGative
*RST Value
EITHer
Examples
TRIG:SEQ2:SLOP:VOLT POS TRIG:ACQ:SLOP:VOLT EITH
Query Syntax
TRIGger:SEQuence2:SLOPe:VOLTage?
TRIGger:ACQuire:SLOPe:VOLTage?
Returned Parameters
<CRD>
Related Commands
TRIG:SEQ2:SLOP:CURR

Table of Contents