Language Dictionary - 4
79
TRIGger:SEQuence2:SOURce
TRIGger:ACQuire:SOURce
Keysight 66312A, 66332A Only
These commands select the trigger source for measurement triggers as follows:
BUS
INTernal
Command Syntax
TRIGger:SEQuence2:SOURce<source>
TRIGger:ACQuire:SOURce<source>
Parameters
BUS | INTernal
*RST Value
INTernal
Examples
TRIG:ACQ:SOUR BUS
Query Syntax
TRIGger:SEQuence2:SOURce?
TRIGger:ACQuire:SOURce?
Returned Parameters
<CRD>
TRIGger:SEQuence1:DEFine
TRIGger:SEQuence2:DEFine
TRIGger:SEQuence2:DEFine applies to Keysight 66312A, 66332A Only
These commands define the names that are aliased to trigger sequences 1 and 2. The command accepts
only ACQuire for sequence 2 and TRANsient for sequence 1 as predefined names. The query allows the
user to query the instrument names aliased to sequences 1 and 2.
Command Syntax
TRIGger:SEQuence1:DEFine TRANsient
TRIGger:SEQuence2:DEFine ACQuire
Parameters
TRANsient, ACQuire
Examples
SEQ1:DEF ACQ SEQ2:DEF TRAN
Query Syntax
TRIGger:SEQuence1:DEFine?
TRIGger:SEQuence2:DEFine?
Returned Parameters
<CRD>
Related Commands
TRIG:SEQ2:ACQ TRIG:SEQ1:TRAN
*TRG
This common command generates a trigger when the trigger subsystem has BUS selected as its source.
The command has the same affect as the Group Execute Trigger (<GET>) command.
In RS-232 mode, this command emmulates some of the functionality of the IEEE-488 Group Execute
Trigger command.
Command Syntax
*TRG
Parameters
None
Related Commands
ABOR INIT TRIG[:IMM] <GET>