Setup
Oscilloscope Setup
1. Set Ch1 vertical scale to 500 mV/div and horizontal scale to 5 ms/div.
2. Set Ch1 Input Impedance to 1M Ohm.
B2961A/62A Setup
1. Connect the banana cables to the B2961A/62A’s Ch1 High/Low Force Terminals
2. Connect the alligator clips to the ends of the 1.1 kΩ resistor.
3. Probe across the 1.1 kΩ resistor to monitor the output using the oscilloscope.
1.1 kΩ
Resistor
U8201A
To oscilloscope’s input
B2961A/62A
05 | Keysight | Sourcing Precise Current/Voltage Sinusoidal Waveforms Using the B2961A/62A – Demo Guide