Series E4360 Service Guide 35
3
Calibration
Calibration Description ............................................................... 36
Calibration Procedure ................................................................. 37
This chapter includes calibration procedures for the Keysight E4360
Series Solar Array Simulator. Instructions are given for performing the
procedures from either the front panel or a controller over the remote
interface.
Refer to the “Equipment Required” section in chapter 2 for a list of the
equipment required for calibration. Also refer to “Measurement
Techniques” for information about connecting the voltmeter and current
shunt.
NOTE
Perform the verification tests before calibrating your solar array
simulator. If the solar array simulator passes the verification tests, the
unit is operating within its calibration limits and does not need to be re-
calibrated.
Calibration Interval
The recommended calibration interval for Keysight E4361A/E4362A
output modules is one year. Keysight E4360A SAS mainframes do not
require calibration.

Other manuals for Keysight Technologies E4360 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight Technologies E4360 Series and is the answer not in the manual?

Keysight Technologies E4360 Series Specifications

General IconGeneral
BrandKeysight Technologies
ModelE4360 Series
CategorySwitch
LanguageEnglish

Related product manuals