Troubleshooting 5
Series E4360 Service Guide 55
Backplane Board Troubleshooting
Troubleshooting the backplane board consists of visually inspecting the
pins and pin connections.
If there are missing or broken pins, replace the board.
If there are bent pins, carefully try to straighten the pin. If you are
unsuccessful, replace the board.
Closely examine the solder connections at the pins. If there are cracks in
the solder, carefully re-solder the connector. If unsuccessful, replace the
board.