Disassembly 4
Series E4360 Service Guide 43
Removing/Installing Output Modules
CAUTION
If you are only installing a single power module, it must be installed in
slot 1. The unused slot must have a filler module installed to ensure
proper cooling.
Step 1. Remove the blower cover. Remove three screws from the top of the
cover and four screws on the sides. Tilt the cover up and slide it out.
Step 2. To remove a module, first remove the two fastening screws at either end
of the module. Grasp the module at the ends and lift it straight up out of
the mainframe.
Step 3. To install a power module, align the module over the alignment pins, and
push it down onto the mainframe connector.
Step 4. Fasten the power module or filler module to the mainframe. Use the
screws provided with the module and install them at either end of the
module.
Step 5. Replace the blower cover when finished. Be careful not to bend the
spring clips on the edge of the cover when sliding it under the lip of the
power modules.