Disassembly 4
Series E4360 Service Guide 45
Accessing the Calibration switch
Step 1. Remove the blower cover. Remove three screws from the top of the
cover and two screws on the sides. Tilt the cover up and slide it out.
Step 2. Remove the interface cover. Remove the three screws along the left side
and two at the rear of the unit. Then remove the screw at the front of the
cover. Lift off the cover.
Step 3. The calibration switch is on the interface board near the ribbon cable. To
change the calibration switch settings, use a small screwdriver to move
the switches. Refer to chapter 3 for settings information.
CAUTION
Do not use a pencil to move the switches. Any graphite dust that gets
on the switches will conduct electricity.
Step 4. Replace all covers when finished.