EasyManuals Logo

Keysight Technologies E4360A User Manual

Keysight Technologies E4360A
118 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #107 background imageLoading...
Page #107 background image
SCPI Commands Appendix C
Series E4360 User’s Guide 107
SCPI Command
Description
:PROTection
[:LEVel] <NRf>, (@chanlist)
Sets the over-voltage protection level
:SAS
:VMP <NRf>, (@chanlist)
Sets the voltage at the peak power point of the curve
:VOC <NRf>, (@chanlist)
Sets the open-circuit voltage
:SLIMit
:HIGH <NRf+> | MAXimum [,(@chanlist)]
Sets the high soft limit for programming the output voltage
:TABle
:OFFSet <NRf> [,(@<chanlist>)]
Adds a voltage offset when operating in Table mode
STATus
:OPERation
[:EVENt]? [(@chanlist)]
Returns the value of the operation event register
:CONDition? [(@chanlist)]
Returns the value of the operation condition register
:ENABle <NRf> [,(@chanlist)]
Enables specific bits in the Event register
:NTRansition <NRf> [,(@chanlist)]
Sets the Negative transition filter
:PTRansition <NRf> [, (@chanlist)]
Sets the Positive transition filter
:PRESet
Presets all enable and transition registers to power-on
:QUEStionable
[:EVENt]? [(@chanlist)]
Returns the value of the questionable event register
:CONDition? [(@chanlist)]
Returns the value of the questionable condition register
:ENABle <NRf> [,(@chanlist)]
Enables specific bits in the Event register
:NTRansition <NRf> [,(@chanlist)]
Sets the Negative transition filter
:PTRansition <NRf> [,(@chanlist)]
Sets the Positive transition filter
SYSTem
:CHANnel
[:COUNt]?
Returns the number of output channels in a mainframe
:MODel? [(@chanlist)]
Returns the model number of the selected channel
:OPTion? [(@chanlist)]
Returns the option installed in the selected channel
:SERial? [(@chanlist)]
Returns the serial number of the selected channel
:COMMunicate
:RLSTate LOCal | REMote | RWLock
Specifies the Remote/Local state of the instrument
:ERRor?
Returns the error number and error string
:GROup
:CATalog?
Returns the groups that have been defined
:DEFine [(@chanlist)]
Group multiple channels together to create a single output
:DELete <channel>
Removes the specified channel from a group
:ALL
Ungroups all channels
:PARallel AUTO | DIRect [, (@chanlist)]
Specifies how output modules are connected in parallel
:MMEMory
:INITialize
Initializes the instrument's memory to the factory-default
settings
:SANitize
Implements the sanitizing standard DoD 5220.22-M
:PASSword
:FPANel
:RESet
Resets the front panel lock password to zero
:REBoot
Returns the unit to its power-on state
:VERSion?
Returns the SCPI version number
TRIGger
[:TRANsient]
[:IMMediate] [(@chanlist)
Triggers the output immediately
:SOURce
BUS|DWEL<n>|PIN<n>|TRAN<n>[,(@chanlist)]
Sets the output trigger source

Other manuals for Keysight Technologies E4360A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight Technologies E4360A and is the answer not in the manual?

Keysight Technologies E4360A Specifications

General IconGeneral
BrandKeysight Technologies
ModelE4360A
CategorySwitch
LanguageEnglish

Related product manuals