Operation and Service Guide 1
Keysight E4412A and E4413A Operating and Service Guide 25
Operation
Operation precautions
If the following energy and power levels are exceeded, the power meter system
may be damaged.
a Maximum Average Power: 200 mW (+23 dBm)
b Maximum Peak Power: 200 mW (+23 dBm)
Maximum torque at the connector should not exceed 12 in-lb (135 Ncm) for the
Type-N connector, or 8 in-lb (90 Ncm) for the 3.5-mm connector to avoid damage
to the connector.
Connect the power sensor by turning only the hex nut portion of the connector.
Damage can occur if torque is applied to the power sensor body.
The connector plastic insulator bead deteriorates when contacted by acetone,
trichloroethylene, carbon tetrachloride, benzene, etc. Refer to Application Note
326, Principals of Microwave Connector Care (5954-1566) or Microwave
Connector Care (08510-90064) for proper cleaning methods.
WARNING
BEFORE CONNECTING THE POWER SENSOR TO OTHER INSTRUMENTS,
ensure that all instruments are connected to the protective (earth) ground.
Any interruption of the protective earth grounding will cause a potential
shock hazard that could result in personal injury and cause damage to the
power sensor.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight Technologies E4412A and is the answer not in the manual?

Keysight Technologies E4412A Specifications

General IconGeneral
BrandKeysight Technologies
ModelE4412A
CategoryAccessories
LanguageEnglish

Related product manuals