A Appendix
46 Keysight E4412A and E4413A Operating and Service Guide
Table A-2 E4412A calibration factor uncertainty at 1 mW (0 dBm) (serial prefixes below
US3848/MYxxxxxxxx/SGxxxxxxxx)
Frequency Uncertainty (%)
10 MHz 2.7
30 MHz 2.7
50 MHz Reference
100 MHz 2.7
1.0 GHz 2.7
2.0 GHz 3.1
4.0 GHz 3.1
6.0 GHz 3.1
8.0 GHz 3.1
10.0 GHz 3.1
11.0 GHz 3.1
12.0 GHz 3.3
14.0 GHz 3.3
16.0 GHz 3.3
18.0 GHz 3.3