Operation and Service Guide 1
Keysight E4412A and E4413A Operating and Service Guide 39
Cleaning
Cleaning solutions
Keeping in mind its flammable nature; a solution of pure isopropyl or ethyl alcohol
can be used to clean the connector.
Connector cleaning
Clean the connector face using a cotton swab dipped in isopropyl alcohol. If the
swab is too big use a round wooden toothpick wrapped in a lint free cotton cloth
dipped in isopropyl alcohol. Refer to Keysight Application Note 326, Principals of
Microwave Connector Care (5954-1566) or Microwave Connector Care
(08510-90064) for proper cleaning methods.
CAUTION
The RF connector beads deteriorate when contacted by hydrocarbon
compounds such as acetone, trichloroethylene, carbon tetrachloride, and
benzene.
CAUTION
Clean the connector only at a static free workstation. Electrostatic discharge
to the center pin of the connector will render the power sensor inoperative.