32 1000 X-Series Oscilloscopes Service Guide
2 Testing Performance
To verify trigger sensitivity
These tests verify the trigger sensitivity. In these tests, you will apply a sine wave
to the oscilloscope at two bandwidths: 10 MHz and the maximum bandwidth of
the oscilloscope. For each sine wave, you will decrease the amplitude of the signal
to a specified level and check to see if the oscilloscope still triggers.
The external trigger sensitivity is also tested in both the 1.6 V and 8 V ranges for
the 70 MHz and 100 MHz (DSOX) models.
Table 9 Equipment Required to Verify Trigger Sensitivity
Equipment Critical Specifications Recommended Model/Part
Signal Generator 10 MHz and 50 MHz, 70 MHz, or
100 MHz sine waves
Keysight N5181A
Power splitter Outputs differ < 0.15 dB Keysight 11667A
Power Meter Keysight N1914A
Power Sensor Keysight E9304A or N8482A
Cable BNC, Qty 3 Keysight 10503A
Adapter N (m) to BNC (f), Qty 3 Keysight 1250-0780
Feedthrough 50Ω BNC (f) to BNC (m) Keysight 0960-0301