EasyManua.ls Logo

Keysight Technologies InfiniiVision 1000 X-Series - Table 10 Internal Trigger Sensitivity Specifications; Test Internal Trigger Sensitivity

Keysight Technologies InfiniiVision 1000 X-Series
60 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Testing Performance 2
1000 X-Series Oscilloscopes Service Guide 33
Test Internal Trigger Sensitivity
Follow these steps to perform the internal trigger sensitivity test:
1 On the oscilloscope, press the [Default Setup] key.
2 Press the [Mode/Coupling] key; then, press the Mode softkey to select Normal.
3 Connect the equipment (see Figure 4).
a Connect the signal generator output to the oscilloscope channel 1 input.
b Connect a 50 ohm feedthrough termination between the channel 1 input
and the BNC cable.
Table 10 Internal Trigger Sensitivity Specifications
Oscilloscope
Models
Internal Trigger Sensitivity Specification Band wid th
<= 10 MHz Maximum Band wid th
50 MHz greater of 0.6 div or 2.5 mV greater of 0.9 div or 3.8 mV
70 MHz greater of 0.6 div or 2.5 mV greater of 0.9 div or 3.8 mV
100 MHz greater of 0.6 div or 2.5 mV greater of 1.2 div or 5 mV

Table of Contents

Other manuals for Keysight Technologies InfiniiVision 1000 X-Series

Related product manuals