36 1000 X-Series Oscilloscopes Service Guide
2 Testing Performance
Test External Trigger Sensitivity
This test applies to all models.
Verify the external trigger sensitivity at these settings:
10 MHz, 250 mV
pp
in 8 V range
10 MHz, 50 mV
pp
in 1.6 V range (DSOX models)
Max BW of oscilloscope, 500 mV
pp
in 8 V range
Max BW of oscilloscope, 100 mV
pp
in 1.6 V range (DSOX models)
1 Connect the equipment (see Figure 5).
a Use the N cable to connect the signal generator to the power splitter input.
b Connect one output of the power splitter to the Ext Trig input through a 50Ω
feedthrough termination.
c Connect the power sensor to the other output of the power splitter.
Table 11 External Trigger Sensitivity Specifications
Oscilloscope
Models
Range
Setting
External Trigger Sensitivity Specification Band width
<= 10 MHz Maximum Bandwidth
50 MHz 8 V 250 mVpp 500 mVpp
70 MHz 8 V 250 mVpp 500 mVpp
1.6 V 50 mVpp 100 mVpp
100 MHz 8 V 250 mVpp 500 mVpp
1.6 V 50 mVpp 100 mVpp

Table of Contents

Other manuals for Keysight Technologies InfiniiVision 1000 X-Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight Technologies InfiniiVision 1000 X-Series and is the answer not in the manual?

Keysight Technologies InfiniiVision 1000 X-Series Specifications

General IconGeneral
BrandKeysight Technologies
ModelInfiniiVision 1000 X-Series
CategoryTest Equipment
LanguageEnglish

Related product manuals