2-26 Keysight M937xA PXIe Service Guide
Tests, Adjustments, and Troubleshooting
System Verification
2-
7. To modify the number of ports to be verified or to change the number of
devices to measure, click on the Configure tab and make the desired
selections.
8. Click Run.
9. Follow the instructions for performing a full calibration or recalling an
existing recent calibration.
10.Follow the instructions for performing the system verification, inserting the
verification devices as prompted.
If the System Fails the Verification Test
1. Disconnect and clean the device that failed the verification test.
2. Reconnect the device, making sure that all connections are torqued to the
proper specifications.
3. Measure the device again.
4. If the analyzer still fails the test, check the measurement calibration by
viewing the error terms as described in Accessing Error Terms” on
page 1-7.
5. Refer to Figure 2-10 on page 2-27 for additional troubleshooting steps.
Inspect all connections. DO NOT remove the cable from the network
analyzer test port. This will invalidate the calibration that you performed
earlier.

Table of Contents

Other manuals for Keysight Technologies M937 A Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight Technologies M937 A Series and is the answer not in the manual?

Keysight Technologies M937 A Series Specifications

General IconGeneral
BrandKeysight Technologies
ModelM937 A Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals