266 N9030B PXA Signal Analyzer Service Guide
LO Synthesizer/Reference Troubleshooting
A16 Reference Assembly Description
A16 Reference Assembly Description
Purpose
The A16 Reference Assembly provides the fundamental reference signals from
which all instrument local oscillator and synchronous timing signals are
derived. The reference board also uses these same signals to generate CW and
modulated RF calibration signals and a 10 MHz time base signal for the
instrument.
The Reference assembly must be replaced as a single assembly, although it is
constructed of two boards, the A16 Reference Main board and A16A1
Reference Daughter board. These boards are listed individually in the Show
Hardware screen, allowing identification of the individual board part numbers
comprising the complete assembly.
The reference signals and local oscillators are phase locked to a timebase in
one of two ways:
1. The internal 10 MHz reference located on the A16 Main board.
2. An externally provided reference such as a house standard or the reference
from another instrument or a base transceiver system. The external
reference can be any RF signal in the range of 1 MHz to 50 MHz, -5 dBm to
+10 dBm.
If an external reference signal is within 200 Hz of the specified external
reference frequency and Freq Ref In is set to Sense, the analyzer will
automatically switch from SENSE:INT to SENSE:EXT.
The phase noise performance of the reference signals establishes the limit on
attainable phase noise performance of the LO signals derived from them.
Therefore, excellent reference signals phase noise is required to achieve
excellent LO signal phase noise.
Status monitoring is comprehensive. Every PLL is monitored for phase lock and
every ALC is monitored for correct leveling. A set of LED's which indicate
whether the 100 MHz VCXO PLL and the 2400 MHz CRO PLL is phase locked
or unlocked. Each oscillator (10 MHz, 100 MHz, 2400 MHz) is monitored for the
presence of RF power (dead/alive status).

Table of Contents

Other manuals for Keysight Technologies N9010B-526

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight Technologies N9010B-526 and is the answer not in the manual?

Keysight Technologies N9010B-526 Specifications

General IconGeneral
BrandKeysight Technologies
ModelN9010B-526
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals