Front Panel Menu Reference SCPI Command
Select Protect\UVP
Enter a value in the UVP level box.
Then check Enable UVP
Enter a value in the UVP Delay box.
Then press Select.
To set the UVP level to 2 volts:
VOLT:PROT:LOW 2
To enable low-voltage protection:
VOLT:PROT:LOW:STAT ON
To specify a protection delay of 5 seconds:
VOLT:PROT:LOW:DEL 5
Clear Output Protection
If an over-voltage, over-current, over-temperature, power-fail condition, power-limit condition,
protection condition, or inhibit signal occurs, the output is disabled. The appropriate operating status
indicator on the front panel will be on. To clear the protection function and restore normal operation,
first remove that condition that caused the protection fault. Then, clear the protection function as
follows:
Front Panel Menu Reference SCPI Command
Select Protect\Clear
Select Clear.
To clear a protection fault:
OUTP:PROT:CLE
The output is re-enabled once the ouptut protection function is cleared.
Keysight RP7900 Series Operating and Service Guide 123
4 Using the Regenerative Power System