Supplemental Characteristics
Supplemental characteristics are not warranted but are descriptions of performance determined
either by design or by type testing. All supplemental characteristics are typical unless otherwise
noted.
Characteristic Keysight SD1000A
Power relay ratings
Option 500:
Option 950:
500 VDC; 60 ADC
950 VDC; 60 ADC
Power relay, mechanical life:
Power relay, hot-switched life:
Sense relay life:
1,000,000 cycles
10,000 cycles
1,000,000 cycles
Discharge resistor
Option 500:
Option 950:
25 kΩ, 10 W
97.5 kΩ, 10 W
Regulatory Compliance
EMC:
Safety:
Complies with European EMC Directive for test and measurement products
Complies with Australian standard and carries C-Tick mark
This ISM device complies with Canadian ICES-001
Cet appareil ISM est conforme à la norme NMB-001 du Canada
Complies with European Low Voltage Directive and carries the CE-marking.
Conforms to US and Canadian safety regulations.
Declarations of Conformity for this product may be downloaded from the Web. Go to
http://regulations.corporate.keysight.com and click on “Declarations of Conformity.”
Environmental
Operating environment:
Temperature range:
1
Relative humidity:
Altitude:
Storage temperature:
Indoor use, installation category II (for AC input), pollution degree 2
0°C to 55°C
95% or less (non-condensing)
Up to 2000 meters
-30°C to 70°C
Output Terminal Isolation: No output terminal may be more than ±950 VDC from any other terminal or chassis ground.
AC Input
Nominal Rating:
Input Range:
Power consumption:
Single phase; 100–240 VAC input, 50-60 Hz
86–264 VAC; 47–63 Hz
150 W
Typical Weight 33 lbs (15 kg)
1
Maximum continuous power available is derated at 1% of rating per degree C from 40°C to 55°C
Appendix A Keysight SD1000A Safety Disconnect System
302 Keysight RP7900 Series Operating and Service Guide

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight Technologies RP7942A and is the answer not in the manual?

Keysight Technologies RP7942A Specifications

General IconGeneral
BrandKeysight Technologies
ModelRP7942A
CategoryPower Supply
LanguageEnglish

Related product manuals