l If the password is set to 0, password protection is removed and the ability to enter calibration
mode is unrestricted. The as-shipped setting is 0 (zero).
l To change the password: unsecure calibration memory with old code, then set the new code.
l If you forget your password, refer to Calibration Switches.
l This setting is non-volatile; it will not be changed by power cycling or *RST.
CALibrate:RESistance:BOUT
Calibrates the bottom out resistance.
Parameter Typical Return
(none) (none)
Calibrates the bottom out resistance: CAL:RES:BOUT
CALibrate:SAVE
Saves the calibration constants in non-volatile memory. Do this at the end of the calibration to avoid
losing changes.
Parameter Typical Return
(none) (none)
Store calibration constants into non-volatile memory: CAL:SAVE
CALibrate:STATe 0|OFF|1|ON [,<password>]
CALibrate:STATe?
Enables or disables calibration mode. Calibration mode must be enabled for the instrument to accept
any calibration commands. The first parameter specifies the state. The second optional parameter is
the password.
Parameter Typical Return
0|OFF|1|ON, *RST OFF 0or1
<password>a numeric value up to 15 digits (none)
Disable calibration: CAL:STAT OFF
Enable calibration: CAL:STAT ON [,value]
A password is required if <password> has been set to a non-zero value.
Keysight RP7900 Series Operating and Service Guide 181
5 SCPI Programming Reference

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight Technologies RP7942A and is the answer not in the manual?

Keysight Technologies RP7942A Specifications

General IconGeneral
BrandKeysight Technologies
ModelRP7942A
CategoryPower Supply
LanguageEnglish

Related product manuals