:FUNCtion
:CURRent 0|OFF|1|ON Enables/disables external current logging.
:MINMax 0|OFF|1|ON Enables/disables external min/max current logging.
:VOLTage 0|OFF|1|ON Enables/disables external voltage logging.
:MINMax 0|OFF|1|ON Enables/disables external min/max voltage logging.
:PERiod <value> Sets the external datalog integration time.
:FUNCtion
:CURRent 0|OFF|1|ON Enables/disables current measurements.
:VOLTage 0|OFF|1|ON Enables/disables voltage measurements.
:SWEep
:NPLCycles <value> Sets the measurement time in number of power line cycles.
:OFFSet
:POINts <value> Defines the offset in a data sweep for triggered measurements.
:POINts <value> Defines the number of points in a measurement.
:TINTerval <value> Defines the time period between measurement samples.
:WHOur
:RESet Resets the accumulated watt-hour measurement.
:WINDow
[:TYPE] HANNing|RECTangular Selects the measurement window.
[SOURce:]
ARB
:COUNt <value>|INFinity Specifies the number of times the Arb repeats.
:CURRent
:CDWell
[:LEVel] <value>{,<value>}|<Block> Specifies the level of each point in the Arb.
:DWELl <value> Specifies the dwell time of each point in the Arb.
:POINts? Returns the number of points in the Arb.
:FUNCtion
:TYPE CURRent|VOLTage Specifies either a voltage or current Arb.
:TERMinate
:LAST 0|OFF|1|ON Selects the output setting after the Arb ends.
[SOURce:]
CURRent
[:LEVel]
[:IMMediate]
[:AMPLitude] <value> Sets the output current when in current priority mode
:TRIGgered
[:AMPLitude] <value> Sets the triggered output current
:BWIDth
:RANGe 0 | 1 Sets the current compensation. (RP793xA,RP794xA)
:LEVel 0 | 1, <value> Sets the compensation frequency. (RP793xA, RP794xA)
:LIMit